Infrarot-Temperaturmessung im Gehörgang mit dem DIATEK 9000 Instatemp und dem DIATEK 9000 Thermoguide Einflußgrößen und Vergleich mit anderen Methoden der Temperaturmessung des Körperkerns (Infrared thermometry in the auditory canal: evaluation on the Diatek 9000 Instatemp and Diatek 9000 Thermoguide)

Nov 1, 1996Der Anaesthesist

Measuring ear canal temperature with two DIATEK devices compared to temperature readings from other body parts

AI simplified

Abstract

Mean differences between infrared thermometry and tympanic probes were -0.41 +/- 0.67 degree C.

  • Shortening the interval between consecutive temperature readings may increase discrepancies, with the second reading typically lower.
  • Positioning infrared thermometers in the auditory canal for 5 seconds resulted in significantly lower recorded temperatures than immediate measurements.
  • Rotating the infrared devices away from the optimal position led to greater disagreement with contact probe measurements.
  • The lower temperatures recorded by infrared thermometry suggest these readings reflect the auditory canal temperature rather than the tympanic membrane.
  • Correcting infrared thermometry results to core-equivalent temperatures minimizes mean differences but does not alter the limits of agreement with traditional methods.