Measuring the Diabetes Drug Semaglutide in Rat Blood and Studying How It Moves Through the Body

Updated

Abstract

The assay achieved a lower limit of quantification of 1 ng/mL with linearity from 1 to 500 ng/mL.

  • The method allows for chromatographic separation in just 9 minutes, significantly faster than previous methods.
  • Intra- and inter-day accuracy ranged from 94.93% to 103.76%, indicating reliable measurements.
  • Precision was maintained at levels of 4.50% or less for intra-day and 5.88% or less for inter-day assessments.
  • Recovery rates were consistent at 93.05-107.95% across quality control levels.
  • Matrix effects remained stable between 96.34-104.12%, suggesting minimal interference from plasma components.