ACS sensors

Single-step CRISPR test for detecting HIV and syphilis directly from whole blood at the point of care

Updated

Abstract

A single-step duplex CRISPR assay can detect both HIV and Treponema pallidum in whole blood within 30 minutes with a limit of detection of 2.5 copies/µL.

  • The assay integrates multiple processes, including reverse transcription and nucleic acid amplification, for simultaneous testing.
  • A low-cost microfluidic chip enables rapid plasma separation and nucleic acid release without the need for external equipment.
  • The system demonstrates 100% sensitivity and specificity when compared to real-time PCR in 51 clinical samples.
  • A smartphone-based portable device facilitates thermal incubation and automated image analysis for user convenience.
  • The approach offers a potential solution for decentralized and efficient pathogen testing.