Se pu = Chinese journal of chromatography

Measuring specific proteins using gel separation and real-time fluorescence imaging

Updated

Abstract

The new protein immune PAGE method can quantify proteins with a linear range from 5.0 to 200.0 mg/L in just 1.5 hours.

  • This method combines immune recognition with online fluorescence imaging for rapid protein quantification.
  • Formaldehyde crosslinking stabilizes antigen-antibody complexes, ensuring reliable results.
  • It shows good accuracy with a maximum relative standard deviation of 1.65% across all measurements.
  • The limit of detection is 0.5 mg/L, with recovery rates between 98.2% and 105.0%.
  • Specificity testing confirms accurate quantification of human transferrin without interference from other proteins.
  • This approach could be adapted for quantifying other proteins using different fluorescently labeled antibodies.

Simplified

Full Text

We can’t show the full text here under this license.

What Lands in Your Inbox Each Week:

  • 📚7 fresh studies
  • 📝plain-language summaries
  • direct links to original studies
  • 🏅top journal indicators
  • 📅weekly delivery
  • 🧘‍♂️always free